Teledyne Pet Fence 201A User Manual

MANUAL ADDENDUM  
MODEL 201A  
AMMONIA ANALYZER  
TELEDYNE ADVANCED POLLUTION INSTRUMENTATION  
(T-API)  
9480 CARROLL PARK DRIVE  
SAN DIEGO, CA 92121-5201  
TOLL-FREE: 800-324-5190  
FAX: 858-657-9816  
TEL: 858-657-9800  
04689  
REV. B1  
Copyright 2005 T-API Inc.  
12 September, 2005  
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
TABLE OF CONTENTS  
2
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
5
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
1.0 M201A AMMONIA ANALYZER  
This manual is to be used in conjunction with the M200A NOx Analyzer manual p/n  
02246 supplied with this instrument.  
The M201A Ammonia Analyzer consists of a modified M200A Nitrogen Oxides Analyzer  
and M501NH Ammonia Converter:  
A functional block diagram of the system is shown in Figure 1.  
The instrument comes in three pneumatic configurations:  
The basic system is shown in Figure 3;  
The M201A with Z/S valves is shown in Figure 4;  
The M201A with External Calibrator Option (ECO) is shown in Figure 5.  
Figure 2 can be used to connect the M201A system for installation. Please note that for a  
M201A with either a Zero/Span or ECO option, you must connect the umbilical cord,  
02255, between the analyzer and the converter assembly.  
1.1 Principle of Operation  
The Teledyne API Model 201A Analyzer measures ammonia by converting NH3 to nitric  
oxide by the following reaction:  
4NH3 + 5O2 4NO + 6H2O  
The resulting NO is then measured by the chemiluminescent reaction of NO with ozone.  
Consult Section 4.1 in the M200A manual (provided with this instrument) for more  
details on NO measurement.  
A molybdenum converter is continuously in the pneumatic pathway, providing the TN  
channel signal. The M501NH ammonia converter is periodically switched so both  
converters are in series, creating the TNx signal.  
The ammonia concentration is computed by the difference between the TNx and TN  
channels, as shown in the following equation:  
NH3 TNx - TN  
See the pneumatic diagram in Figure 1.  
The actual formula for computation of ammonia concentration is more complicated than  
the above equation, as it includes converter efficiencies, plus sample temperature and  
pressure. The difficulty in ammonia measurement is that the molybdenum converter  
6
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
converts some of the ammonia to NO, thus the TN channel data contains a small  
ammonia signal. Secondly, the ammonia converter does not convert all of the ammonia  
or ambient NO2 to NO. Finally, the high temperature in the ammonia converter causes  
the oxygen and nitrogen in ambient air to react, creating a background of NOx which  
needs to be factored into computing the result. Fortunately, all of the above  
shortcomings have been overcome in the M201A by careful calibration, and calculation  
of the result. .  
Periodically, the AutoZero valve switches allowing the analyzer to read zero  
background. The AutoZero readings are subtracted from all of the other readings. This  
improves zero baseline stability.  
As with many chemical reactions the conversion of ammonia in the presence of other  
oxides of nitrogen is complicated.  
In the molybdenum converter operating at 315oC, the following significant reactions are  
taking place:  
Mo + NO2 MoO3 + NO ~100% Eff  
NH3 + Mo + O2 MoO3 + NO <10% Eff  
Notice in the second reaction some of the ammonia is converted to NO by the  
molybdenum metal.  
The M501NH NH3 Converter operates at 825oC. At this high temperature, several  
reactions occur:  
NO NO  
Efficiency = ~100 %  
Efficiency = ~85%  
Efficiency = ~85%  
Efficiency = <1%  
NOx NO  
NH3 NO  
N2 + O2 NO  
As can be seen from the above reactions, the calculation of the ammonia concentration  
and overall calibration of the instrument must be done carefully if accurate ammonia  
concentrations are to be measured. The calibration procedure is covered in Section 2.  
1.1.1 Special Considerations for Ammonia Measurement  
Ammonia is a difficult gas to measure due to its chemical characteristics. Because of  
its large dipole moment, the gas tends to adsorb on surfaces and diffuse into many  
materials. The following precautions should be observed when designing an ammonia  
sampling system and connecting the M201A analyzer to the system:  
Materials:  
Do NOT use TEFLON tubing or fittings in contact with NH3 sample or span gas.  
7
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Do NOT use copper tubing or fittings designed for household plumbing.  
Use ONLY Chromatography grade (cleaned, passivated) stainless steel tubing  
Use ONLY Glass tubing for sample inlet manifold.  
These rules apply also to your calibrator’s internal plumbing. Verify that the calibrator  
pneumatic pathway that contacts ammonia only uses stainless steel or glass tubing.  
The M702SS calibrator supplied with the External Calibration Option is built with  
stainless steel fittings and tubing for all internal plumbing wetted by NH3.  
Typical ambient NH3 levels are < 15 ppb, unless there is some large source of  
ammonia nearby, so the sampling and analysis system must be carefully designed and  
well maintained.  
1.1.2 Sample Filtration  
The M201A does not come with a sample filter installed. Depending on the options  
ordered, a sample filter may be separately packaged with the instrument. If the filter is  
installed, the instrument response time is significantly increased, due to its large surface  
area. Also, the condition of the filter (particulate loading and type) can adversely affect  
the response time.  
For minimum response time operation, the instrument should be supplied with clean  
sample gas at ambient pressure.  
If the sample filter is not included, the only additional filter in the M201A pneumatic  
pathway is a ¼” dia stainless steel sintered filter used to protect the flow control orifice  
from plugging. This filter will plug rapidly if the sample gas is not filtered before entering  
the analyzer. Some means of filtering the sample gas, such as an external stainless  
steel filter should be added upstream of the analyzer.  
8
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
1.2 M201A Analyzer Specifications  
Ranges  
0-50 to 0-2000 ppb in 1 ppb increments  
Single, Independent, AutoRange  
0.5 ppb RMS  
Range Modes  
Noise at Zero1  
Noise at Span1  
< 1.0% of reading above 50 ppb  
Lower Detectable Limit 1 ppb RMS  
Zero Drift3  
2 ppb / 24 hours  
Span Drift  
< 1.0% FS Range / 7 days  
Lag Time  
40 sec  
Rise Time  
90% 120 sec  
Fall Time  
90% 120 sec  
Sample Flow Rate  
Linearity  
1000 cc/min  
2% of full scale  
Precision  
1.0% of reading  
Temp Range  
Dimensions HxWxD  
Weight, Analyzer  
Weight, Converter  
Weight, Pump  
Power, Analyzer  
10-30oC  
7” x 17” x 23.6” (18cm x 43cm x 61cm)  
43 lbs (20 kg)  
24 lbs (11 kg)  
16 lbs (7 kg)  
100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz,  
125 watts  
230V ~50 Hz, 125 watts  
Power, Analyzer5  
Power, Pump  
110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts  
Power, Pump CEMark5 230 V ~50 Hz, 2.5 A peak  
Environmental  
Installation Category Pollution Degree 2, Over-voltage  
Category II  
Analog Output  
0-100, 0-1, 0-5, 0-10VDC bi-polar, 4-20mA isolated  
Analog Resolution  
Status Option  
Measurement Units  
1 part in 2048  
12 Status Outputs  
ppb, ug/m3  
1. As defined by USEPA  
2. Defined as twice the zero noise level  
3. At constant temperature and voltage  
4. Bi-polar  
5. Electrical rating for CE Mark Compliance  
9
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
2.0 CALIBRATION PROCEDURE  
Calibration of the M201A is done by first calibrating the TN and TNx channels with NO  
span gas diluted with zero air. In the second phase of calibration, ammonia span gas is  
introduced to measure the overall system efficiency for NH3 conversion. Unlike NOx  
analysis, the NH3 converter efficiency must be measured because it is considerably  
lower than 100% and may change as the converter ages.  
If the instrument is configured as a basic model, all zero/span gasses are introduced  
through the sample port. If the Zero/Span valves option or the ECO option is ordered,  
the zero/span gas is introduced through the zero/span valves. There are several ways  
of automating the calibration or span check process, they are covered in Section 7 of  
the M200A manual. After reviewing the procedure described below, it should be simple  
to adapt the procedures in Section 7 to the M201A, if necessary.  
2.1 Zero/Span Calibration – Basic Instrument  
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the  
expected zero value.  
10  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Table 2.1 - Zero Calibration Procedure - Zero Gas Through the Sample Port  
Step No.  
Action  
Comment  
1.  
Press  
CAL  
The M201A enters the calibrate mode from  
sample mode. Input zero gas through the  
sample port.  
2.  
3.  
Wait 20  
min  
Wait for reading to stabilize at zero value.  
Press  
ZERO  
If you change your mind after pressing ZERO,  
you can still press EXIT here without zeroing  
the instrument.  
4.  
5.  
Press  
ENTR  
Pressing ENTR actually changes the  
calculation equations.  
Press  
EXIT  
M201A returns to sampling. Immediately after  
calibration, data is not added to the DAS  
averages.  
Table 2.2 - Enter the Expected Span Gas Concentrations Procedure  
Step No.  
Action  
Press  
Comment  
This key sequence causes the M201A to  
prompt for the expected TNX concentration.  
Enter the TNX span concentration value by  
pressing the key under each digit until the  
expected value is set. This menu can also be  
entered from CALS or CALZ.  
1.  
CAL-CONC-TNX  
2.  
3.  
Press ENTR  
Press  
ENTR stores the expected TNX span value.  
Now enter the expected TN span concentration  
as in step one.  
CAL-CONC-TN  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR stores the TN span value and  
returns the prompt to the CONC menu.  
Returns instrument to SAMPLE mode.  
11  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Table 2.3 - Span Calibration Procedure - Span Gas thru Sample Port  
Step No.  
Action  
Press CAL  
Comment  
1.  
2.  
The M201A enters the calibrate mode. Input NO span  
gas through the sample port.  
Press  
Set the converter efficiency parameter to 1.0000, then  
press ENTR-EXIT.  
CAL-CONV-SET  
Wait 20 min  
Press SPAN  
2.  
3.  
Wait for reading to stabilize at span value.  
If you change your mind after pressing SPAN, you can  
still press EXIT here without spanning the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation  
equations and causes the instrument to read the TN  
and TNx span concentrations.  
M201A returns to sampling. Immediately after  
calibration, data is not added to the DAS averages.  
12  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
2.1.1 Zero/Span Calibration with Z/S Valves or ECO Option  
The zero gas enters through the zero gas inlet port on the rear panel. Make sure that  
the umbilical cord that makes the control connections between the analyzer and the  
converter is connected (see Figure 2 below). The zero gas valve can be activated by a  
variety of methods discussed in Section 7 of the M200A manual. The example below  
will use the front panel keyboard to activate the valves.  
Table 2.4 - Zero Calibration Procedure - Zero Gas from rear panel zero air inlet  
Step No.  
Action  
Comment  
1.  
2.  
Press  
CALZ  
The M201A enters the calibrate mode from  
sample mode. The zero gas is supplied  
through the zero air inlet port on the rear panel.  
Wait 20  
min  
Wait for reading to stabilize at zero. If the  
instrument has recently been exposed to high  
levels of NH3, you may need to wait up to 2  
hours for the instrument to return to zero.  
3.  
Press  
ZERO  
If you change your mind after pressing ZERO,  
you can still press EXIT here without zeroing  
the instrument.  
4.  
5.  
Press  
ENTR  
Pressing ENTR actually changes the  
calculation equations.  
Press  
EXIT  
M201A returns to sampling. This switches the  
ZS valves so that sample gas is coming in the  
sample port on the rear of the instrument.  
Immediately after calibration, data is not added  
to the DAS averages.  
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the  
expected zero value.  
The expected value of the span gas is entered into the instrument by the following  
procedure.  
13  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Table 2.5 - Enter the Expected Span Gas Concentrations Procedure  
Step No.  
Action  
Press  
Comment  
This key sequence causes the M201A to  
prompt for the expected TNX concentration.  
Enter the TNX span concentration value by  
pressing the key under each digit until the  
expected value is set. This menu can also be  
entered from CALS or CALZ.  
1.  
CAL-  
CONC-  
TNX  
2.  
3.  
Press  
ENTR  
ENTR stores the expected TNX span value.  
Press  
Now enter the expected TN span concentration  
as in step one.  
CAL-  
CONC-TN  
4.  
5.  
Press  
ENTR  
Pressing ENTR stores the TN span value and  
returns the prompt to the CONC menu.  
Press  
EXIT  
Returns instrument to SAMPLE mode.  
14  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Table 2.6 - CALS - Span Cal Procedure - Gas from rear panel span gas inlet  
Step No.  
Action  
Press  
Comment  
1.  
The M201A enters the calibrate mode. The  
Z/S valves are activated to route span gas into  
the analyzer.  
CALS  
2.  
Press  
Set the converter efficiency parameter to 1.0000, then  
press ENTR-EXIT.  
CAL-CONV-SET  
2.  
3.  
Wait 20  
min  
Wait for the reading to stabilize at span value.  
Press  
SPAN  
If you change your mind after pressing SPAN,  
you can still press EXIT here without spanning  
the instrument.  
4.  
5.  
Press  
ENTR  
Pressing ENTR actually changes the  
calculation equations and causes the  
instrument to read the TN and TNx span  
concentrations.  
Press EXIT  
M201A returns to sampling. Immediately after  
calibration, data is not added to the DAS  
averages.  
2.1.2 TN/TNx Channel Calibration Quality  
After the TN and TNx channel calibration it is important to check the slopes and offsets  
to make sure they fall within the acceptable range. The tolerances are different than a  
typical NOx analyzer primarily because of the performance of the M501NH ammonia  
converter.  
Slopes can vary from 2.0 to 0.7, and should be within 20% of each other with the TNx  
slope typically smaller than the TN slope.  
Offsets should be zero ±250 mV and be within 100mV of each other.  
If the slope and offsets are within specifications, proceed on the ammonia channel  
calibration.  
15  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
2.1.3 NH3 Calibration  
Since the system efficiency is less than 100%, it is necessary to determine a conversion  
efficiency factor for ammonia and enter it into the system.  
1. Set the converter efficiency to 1.0000 by:  
Press CAL-CONC-CONV-SET on the keyboard  
Press the keys under the numbers until 1.0000 appears.  
Press EXIT to return to the sample mode.  
2. Introduce ammonia span gas of known concentration into the instrument. Allow time for  
instrument stabilization. The final readings will be less than the concentration calculated  
from flows. For example with 400 ppb NH3 span gas:  
TN = 40 ppb  
TNx = 380 ppb  
The 40 ppb TN reading is due to the fact that the moly converts a small amount of the  
NH3 gas to NO. The 380 ppb TNx reading is due to ammonia conversion plus some  
conversion of nitrogen and oxygen in the dilution air.  
3. Continuing with the example above, the instrument will report a NH3 concentration value of  
340 ppb. Divide the reported NH3 concentration by the concentration computed from flow  
data.  
NH3 converter eff. = [NH3M201A] / [NH3FLOWS  
From the above example :  
]
M201A NH3 reading = 340 ppb  
NH3 conc from flows = 400 ppb  
340 / 400 = 0.85 NH3 converter system efficiency  
4. This value now is entered into the instrument.  
Press CAL-CONC-CONV-SET  
Press the buttons under the numbers until the desired value is entered, in this case  
0.8500, then press ENTR.  
Press EXIT to return to SAMPLE mode.  
16  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
The instrument is now fully calibrated.  
2.1.4 Channel Calibration Quality  
After the ammonia channel is calibrated via the converter efficiency parameter, it is  
important to check the calibration by sampling the NH3 calibration gas and confirming  
that the instrument responds with the correct concentration. In the above example the  
M201A should report 400 ppb NH3 concentration when measuring the 400 ppb NH3  
span gas used.  
17  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
3.0 MAINTENANCE SCHEDULE  
3.1 M201A Maintenance Schedule  
The maintenance requirements of the M201A are essentially like a standard M200A  
NOx analyzer. Please refer to Table 8-1 in the M200A Operator Manual for the M200A  
maintenance schedule.  
In Table 3.1 below, there are the extra maintenance items required by the M201A.  
Table 3.1: Preventative Maintenance Schedule  
Item  
Maintenance Interval  
Reference Section  
M501NH3 Converter  
Reaction Cell  
Check every 6 months  
Section 3.2  
Clean quarterly as  
necessary  
Sect 9.3.8, Fig 8-3 in  
the M200A Manual  
AutoZero flow check  
Quarterly as needed  
Section 3.3  
3.2 Maintaining the M501NH Converter  
The M501NH Converter is operated at very high temperature. The center tube in the  
converter slowly oxidizes and requires replacement. The most obvious indicator of  
failure is no ammonia conversion, despite the converter being at temperature. This is  
caused by the oxidized tube breaking off in the converter.  
The following procedure describes how to disassemble and replace converter parts.  
The various parts in the converter become delicate and brittle after prolonged exposure  
to high temperatures. It is therefore a good idea to have a complete set of replacement  
parts on hand before starting, as listed in Table 3.1.  
18  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Table 3.2 – M501NH Converter Rebuild Parts List  
Part No.  
KIT000191  
KIT000139  
Description  
Converter Rebuild Kit  
Ceramic Bushings  
HE0000007 Ceramic Heater 220W 60 VAC  
Procedure:  
1. Turn off power to the converter and allow to cool. IT IS VERY IMPORTANT THAT  
THE CONVERTER BE COOL BEFORE ATTEMPTING ANY DISASSEMBLY OR  
REPAIRS.  
CAUTION VERY HOT  
Will cause severe burns – disassembly while hot will  
damage other converter components. Allow sufficient  
time to cool.  
2. Remove the chassis cover of the M501NH, remove the aluminum cover over the  
oven, remove the two U-shaped hold down clamps.  
3. Disconnect the 1 stainless steel and 2 PTFE tubes from the converter cartridge.  
4. Loosen the nut holding the cartridge assembly in the U-shaped angle bracket  
mounted on the chassis floor.  
5. Gently lift and tilt the converter assembly out of the U-shaped bracket and slide the  
assembly out of the oven.  
6. Loosen the ¼” SS tube fitting nearest the hold-down bracket, this fitting holds the  
central tube in the converter. Replace the tube that is part of KIT000191  
7. Re-assemble the converter by doing the above steps in reverse order.  
19  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
3.3 AutoZero Flow Check  
Since the M201A is an ammonia analyzer, the flow through the AutoZero orifice is  
especially important. Check the AutoZero flow as follows:  
1. This procedure should be performed with the sample pump running.  
2. Fold down the rear panel on the M201A chassis. Locate the vacuum manifold at the  
center rear of the chassis. Locate the 1/8” tube fitting 2nd from the left. It will have a  
label of “0.010”, indicating the flow orifice installed.  
3. Remove the fitting and attach a calibrated flowmeter capable of measuring in the  
range of 500 cc/min. The flowmeter should indicate a flow of 500 cc/min ±10%.  
4. If the flow is outside this range (most likely lower, due to plugging):  
1. Turn off the sample pump and instrument.  
2. Remove the 1/8” pipe-to-tube fitting and remove the ¼” sintered filter (p/n  
FL0001). This filter is meant to protect the orifice, it will usually become plugged and  
need replacement rather than the orifice. Replace the filter, then re-assemble the  
manifold. Re-start the sample pump and re-check the flow.  
20  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Figure 1 - M201A Ammonia Analyzer – Block Diagram  
21  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
NH3 SPAN GAS  
NO SPAN GAS  
CYL  
IN  
1
CYL  
IN  
2
50 PSIG MAX  
INLET PRESSURE  
ZERO AIR IN  
FROM  
ZERO  
AIR  
MODULE  
CAL GAS OUT  
SS TUBING  
SS TUBING  
EXHAUST  
OUT  
TN  
OUT  
SAMPLE  
IN  
ZERO  
AIR  
SS TUBING  
TNX  
OUT  
SPAN  
GAS  
UMBILICAL  
POWER  
SS TUBING  
VENT  
TO VACCUM PUMP  
TO  
CONV  
FROM  
CONV  
UMBILICAL  
02522 UMBILICAL CORD  
Figure 2 - M201A External Pneumatic Connection Diagram with ECO (External Calibrator Option)  
22  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Figure 3 - M201A Pneumatic Diagram Basic  
23  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Figure 4 - M201A Pneumatic Diagram Z/S Option  
24  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Figure 5 - M201A Pneumatic Diagram External Calibrator Option (ECO)  
25  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
Figure 6 – M501NH Converter Cartridge  
26  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
27  
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
4.0 M201A SPARE PARTS LIST  
000940400 Orifice, 4 mil, 80 cc, Rx Cell  
000940600 Orifice, 10 mil, 500 cc, Rx Cell  
002270100 Gasket (Rx Cell) Qty. 12  
002730000 Window 665 NM  
002761020 CPU Board (201A AMX))  
003290000 Thermistor Assembly  
004020200 Flow/Pressure Sensor Board  
005140300 V/F Board  
007040000 Keyboard  
007280000 NEW Display  
010680100 Heater, MOLY Converter  
010860000 Status/Temperature Board  
011310000 Drier Assembly Complete with Flow Control  
011930000 PMT, NOx  
011980000 Assembly, MOLY Thermocouple (Type J)  
012360000 Fan, Power Supply Module  
013140000 Fan, PMT Cooler  
013570000 Thermistor Assembly (Cooler)  
013600000 Operators Manual for M200A  
014020300 M200A 37 mm 1 um Filter Expendables Kit  
014020400 M200A 47 mm 1 um Filter Expendables Kit  
014040000 M200A Level 1 Spare Parts Kit (for 10 units)  
014040100 M200A Spares Kit for 1 Unit  
014080100 Assembly, High Voltage Power Supply  
014610000 Cooler Assembly  
28  
 
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
M201A Spare Parts List (cont.)  
016810200  
017230000  
018720000  
021070000  
022300000  
024710000  
024720000  
024730000  
024750000  
046890000  
046570100  
O3 Generator Assembly for M200A/200AH Low Output  
Dryer Material, Drierite  
Welded Moly Conv. Assy, w/ O3 Scrubber, w/ valves  
PMT Pre-amplifier Board Assembly  
DC Power Supply Board  
Tubing: 6’, 1/8” CLR  
Tubing: 6’, 1/8” BLK  
Tubing: 6’, 1/4” BLK  
Tubing: 6’, 1/4” TYGON  
M201A Manual Addendum  
Catalytic Converter Insert  
CB0000001 FUSE O3, 1A  
CP0000014 Fuji Temperature Controller  
DR0000002 PMT Desiccant Baggies  
FL0000001  
FL0000003  
Sintered Filter  
Filter, DFU  
FM0000004 Flow Meter, 0-1000 cc  
HE0000007 Ceramic Heater 220W 60 VAC  
HE0000017 Heater, Reaction Cell, 12W  
HE0000019 Heater, MOLY Converter  
HW0000020 Spring, Flow Control  
HW0000036 TFE Thread Tape (48 FT)  
HW0000150 Clamp, Hose, Nylon, 1/4”  
KIT000019  
KIT000021  
KIT000036  
KIT000041  
KIT000051  
KIT000058  
KIT000059  
KIT000067  
KIT000103  
KIT000110  
Replacement Cooler Assembly, M100A/M200A  
Replacement PWR SW BD, 'A', M100A/200A  
Retrofit, O3 Killer  
Retrofit, Moly Valves, M200A  
Rebuild Kit, RX Cell Ambient, M200/A/251/252  
Leak Checker with Gauge  
Leak Checker without Gauge  
Retrofit, M200A Rcell Rebuild Kit  
Replacement, Welded Moly Cart, Type J Tc, Heater  
Repl, Welded Moly Type J Tc w/o valves, w/o O3 killer, in can  
29  
 
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1  
M201A Spare Parts List (cont.)  
KIT000129 Welded Moly, Cartridge only  
KIT000191 Rebuild, NH3 INNER TUBE & CATALYST  
OR0000001 O-Ring, Flow Control  
OR0000002 O-Ring, Bearing, Cell  
OR0000010 O-Ring, S, O3 Generator  
OR0000012 O-Ring, Permeation Oven-Units with IZS  
OR0000014 O-Ring, Permeation Oven  
OR0000021 O-Ring, Scrubber  
OR0000025 O-Ring, Zero Air Scrubber  
OR0000034 O-Ring, Input/Output Mirror/Detector  
OR0000042 O-Ring, Sensor Assembly  
OR0000044 O-Ring, Reaction Cell  
OR0000045 O-Ring, Sample Filter  
PS0000010 15V Switching Power Supply  
PU0000005 Pump 1150V/60Hz  
PU0000006 Pump 220V/50Hz  
PU0000011 607 Pump Rebuild Kit  
RL0000003 Solid State Relay, 12 Vdc  
RL0000007 3 AMP Opto Relay  
RL0000008 Solid State Relay, 12 VDC  
RL0000015 Solid State Relay, 115 Vac  
SW0000006 Overheat SW, Cell/Oven  
SW0000008 Pressure Sensor  
VA0000007 Solenoid Valve, Stainless Steel, 12V  
VA0000024 Manifold Valve, 3-Way, Vent  
30  
 

Toshiba Computer Drive SD M1711 User Manual
Toshiba CRT Television 27AF44 User Manual
Triarch Indoor Furnishings 29583 User Manual
Tripp Lite Switch P781 006 User Manual
Uniden GPS Receiver TRAX350 User Manual
Vector Work Light VEC190POB User Manual
ViewSonic Car Video System VS15560 User Manual
Viper GPS Receiver 208 User Manual
Weber Gas Grill 54381 User Manual
Weed Eater Lawn Mower WE12538K User Manual